6
CQM Cable Differences
– Two CQM cables are provided in the standard accessory kit.
The only difference between them is that one cable has a small pin and the other does
not. The cable with the pin is used to enable the REM circuitry on the DUT. The cable
without the pin is used to bypass the REM circuit and allow for shorting of the REM leads.
Note that some generators do not allow for the REM circuit to be disabled. For these
generators it does not matter which cable is used.
When using the ESU-2400 CQM test mode, the cable with the pin must be used.
For other modes, the Dispersive ports on the ESU-2400 are shorted together and the
cable without the pin should be used.
Cable Without Pin,
Disabled REM
Cable With Center
Pin, Enables REM