12
Tissue Test
–
This mode evaluates the tissue response function of the DUT. In this
mode, the ESU-2400 is configured for an external TRL-2420 lamp load. The ESU-2400
is connected in series with the load as shown below. The terminals are labeled Return
and Active for reference only. The DUT Return terminal shown below is typically not the
Patient Return or REM electrode. Refer to the DUT service manual to determine which
DUT ports are used for the tissue response output.
If the output terminals of the DUT are not compatible with the standard ESU-2400 safety
test leads, the alligator clips supplied with the ESU-2400 accessory kit can be used to
connect directly to the electrosurgical instrument as shown in the connection diagram
below.
ESU Generator
(DUT)
Active
Return