173
ESU ANALYZER MODES
RF MEASUREMENT
MODE
Allows use as a general purpose RF meter, including current,
voltage, power and timing measurement
LEAKAGE TEST 1A
Tests the open circuit leakage of an Isolated Type CF
generator. The test complies with IEC 601.2.2, sec. 19.101b,
fig, 104 and sec. 19.102, adopted by ANSI/AAMI HF18-2001
LEAKAGE TEST 1B
Tests the open circuit leakage of an Isolated Type CF
generator. The test complies with IEC 601.2.2, sec. 19.101b,
fig, 104 and sec. 19.102, adopted by ANSI/AAMI HF18-2001
LEAKAGE TEST 2
Tests the leakage to ground of Earth-referenced Type BF
generators. This test complies with IEC 601.2.2, sec.
19.101a, test 1, fig. 102, and sec. 19.102, adopted by
ANSI/AAMI HF18-2001
LEAKAGE TEST 3
Tests the leakage to ground of Earth-Referenced Type BF
generators. This test complies with IEC 601.2.2, sec.
19.101a, test 2, fig. 102, and sec. 19.102, adopted by
ANSI/AAMI HF18-2001
LOAD CURVE
Allows user to automatically generate a power curve based
on varying load, commonly seen in the generator data sheets
REM/ARM/CQM
MODE
Uses internal load bank as a variable resistor to test the
Generator's CQM/RECQM circuit
AUTOSEQUENCE
MODE
Allows user to follow a standard or custom defined protocol
to sequentially buildup a test record, applicable to many
manufacturer's recommended verification and calibration
procedures
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...
Страница 189: ......