121
Networking
– The networking screen shows the current network information and allows
the user to select DHCP (automatic) or Static IP address mode.
Address Type
–
The user can select whether the IP address is received automatically
from a DHCP server or configured for a Static IP address.
NOTE: After changing the
Address Type, the system will need to be rebooted for the changes to take effect.
IP Address
– This field identifies the IP address when DHCP is enabled or allows the
user to enter an IP address when the Static Address Type is selected.
Subnet Mask
– This field identifies the Subnet Mask when DHCP is enabled or allows the
user to enter a Subnet Mask when the Static Address Type is selected.
Default Gateway
– This field identifies the Default Gateway when DHCP is enabled or
allows the user to enter a Gateway address when the Static Address Type is selected.
MAC Address
– This read only field identifies the MAC address of the ESU-2400.
Network Name
– This identifies the ESU-2400 on the network. The default name is
ESU2400_xxxx where “xxxx” is the ESU-2400 serial number. Network names must be
unique for each device on a network. The network name is used when browsing the ESU-
2400 on the network. For the example above, the ESU-2400 could be viewed in Windows
Explorer by typing in \\ESU2400_1011 in the address field.
NOTE: for ESU-2400H model analyzers the network address will be \\ESU2400H_xxxx
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...
Страница 189: ......