26
6. This will open the Pulsed Mode Configuration screen. Select Pulsed Mode 1 by
clicking on the mode button or the picture.
7. A threshold should be chosen so that the pulse will be over the threshold but not
too close to the maximum value to account for variations in the waveform. In this
scenario 600mA is sufficient to detect the 700mA waveform. See pg 89 for more
information on choosing a threshold.
8. Press the button next to Threshold 1 and enter 600 mA for the threshold.
9. Click the Accept Changes button to enter your settings and return to the
measurement screen.
10. Press the Display Parameters button and select Two Readings from the Select
Screen menu.