93
The above image shows typical measurements when “All Readings” is selected from
“Display Parameters”. The voltage, current, power, and time for each “pulse” is displayed,
as well as the duty cycle for “pulse 1”.
Pulse Mode 3
: Pulsed Mode 3 is used for waveforms that have three different amplitudes
to be measured. Below is an example of such a waveform.
Pulse Mode 3 uses two user-selectable thresholds to determine if the waveform is
current
ly in “pulse 1”, “pulse 2”, or “pulse 3”. When the waveform amplitude is greater
than threshold 1 the waveform is considered
to be “pulse 1”. When it is below threshold
1 but above threshold 2 it is considered to be “pulse 2”. When the waveform is below
threshold 2 it is considered to be in “pulse 3”.