DUT Setup Tab
–
The DUT can be triggered by the Operator or the ESU-2400, as shown
below. When Operator is selected, the user will be prompted when to activate or
deactivate the ESU.
When ESU-2400 is selected as the trigger source, the activation can be setup manually
or by DUT manufacturer and model. For Manual Setup, the user selects the desired
footswitch output to be used. If RS-232 is enabled, the user also selects the baud rate
and enters commands to be used for DUT setup and control. When entering commands
use “%P” to represent the power level. In the example below, ESU-2400 would send the
command “:FTP300” when setting the power level to 300 watts.