20
Scenario 2
– I need to measure mA and Watts of a Bipolar output with a 300 ohm
Load. I want the ESU-2400 to trigger the DUT with the footswitch output.
Solution
–
1. Select Measure RF Energy from the main menu.
2. Connect one Bipolar output from the DUT to the Active input on the ESU-2400. If
you are using the cables from the ESU-2400 accessory kit, use the Yellow Lead.
3. Connect the other Bipolar output port from the DUT to the Dispersive 1 input on
the ESU-2400. If you are using the cables from the ESU-2400 accessory kit, use
the Blue cable with banana jacks on each end, not the CQM/RECQM cable.
4. Press the Test Load button to open the load configuration box. The screen
should look like the following:
5. Press the Button next to Internal Load and enter 300 in the number pad. The
load configuration box should automatically close.