9. MAINTENANCE AND TROUBLESHOOTING
Page 9/3
075D0002-01
Open the instrument cover door and press any key. The following menu is displayed:
Using a 5 mL syringe, pour 3 mL of
sysCLEAR
into the WBC chamber and press any key. The
next menu is displayed :
Pour 3 mL of
sysCLEAR
inside the RBC chamber and press any key. The next menu is displayed:
Close the instrument cover door and press any key. The following menu is displayed:
Stars are displayed at the beginning of the cycle. Every 10 seconds a star is cleared off. The
procedure involves different cycles, backflush, aspiration, rinsing, which allow a perfect cleaning
of the apertures. After this procedure is completed, run a startup cycle, then sample analysis can
begin.
CONCENTRATED CLEANING WAIT 2MNS 27S
************
PLEASE CLOSE COVER DOOR
PRESS A KEY TO CONTINUE ...
POUR 3ML OF SYSCLEAR IN RBC CHAMBER
PRESS A KEY TO CONTINUE ...
POUR 3ML OF SYSCLEAR IN WBC CHAMBER
PRESS A KEY TO CONTINUE ...
9.1.3.4. Concentrated cleaning
This procedure provides a strong cleaning of the RBC and WBC apertures. This cycle lasts
approximately 3 minutes and involves the operator intervention to fill the chambers with
sysCLEAR
or diluted bleach solution.
Move the cursor to the function
4
CONCENTRATED CLEANING and press
ENTER
. The following
menu is displayed:
PLEASE OPEN COVER DOOR
PRESS A KEY TO CONTINUE ...
The diluted bleach solution is obtained by a 1:5 dilution of a commercial bleach solu-
tion containing from 10 to 15% of sodium hypochlorite.
NOTE:
WARNING: Commercial strength bleach is 10 to 15% sodium hypochlorite. When handling
bleach, which can be used as a cleaning and antiviral agent, wear protective clothing,
gloves, and safety glasses. It is harmful if swallowed and may cause eye or skin irritation.
Use bleach that is free of heavy metals.
To prepare a 1:5 dilution of a commercial bleach solution, dilute one part of bleach
with five parts of clean distilled water, or clean deionized water. The prepared solution
is stable for one week when stored at room temperature.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...