6. STARTUP AND SAMPLE RUN
Page 6/5
075D0002-01
6.3.3. Standard identification mode
Press the
ID/SEQ
key to enter the control Run #. The following menu is displayed:
Enter the run number (from 1 to 9999) using the numeric keys of the front panel. Press the
ENTER
key to record the number (or the
ESC
key to save the displayed one) : A message "PLEASE
CLOSE HOLDER DOOR" is displayed.
RUN #? :
EXIT : ESC
NEXT :
SAVE : ENTER
Diag. 6.2
The tube holder is associated with 3 switches which
are able to detect the sampling position, the presence
of the tube holder and a wrong position of the tube
holder. The 4 positions can be used for the following
sample tubes (list is not exhaustive):
Position 1 : Vacutainers
Position 2 : Mini Vacutainers
Position 3 : Control and calibration vial position
Position 4 : Micro sample collection devices.
The sample tube holder has 4 positions according to
the sample tube characteristics. The required posi-
tion is selected when it is facing the inside of the
sampling compartment. Turn the sample tube holder
in either side to place the required position until the
'click' is heard.
6.3.4. Sample tube holder selection
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...