7. CALIBRATION AND QUALITY CONTROL
Page 7/8
075D0002-01
B - Calibration failed:
If the statistical figures are not within the acceptable limits, the following menu is displayed:
- Coefficient of variation is not within the limits given in Table 7.1 or,
- The percentage difference between the target and the mean value is greater than 20.
The calibration failed and the results are printed out. The calculated coefficients are rejected, the
instrument keeps in memory the previous calibration coefficients and returns to the calibration
menu.
CALIBRATION
DATE : 04/02/98
TIME : 09:21
CALIBRATION FAILED
OPERATOR : JMG
LOT Nr : SETpoint
RUN
WBC
RBC
HGB
HCT
PLT
1 P
6.3
5.30
16.0
47.2
230
2
6.3
5.35
16.1
47.1
224
3
6.4
5.40
16.2
47
230
4
6.5
5.31
16.1
46.8
240
5
8.2
5.28
16.1
47.5
228
6
6.5
5.30
16.0
47.1
235
MEAN
6.78
5.32
16.1
47.1
231.4
TARGET
6.5
5.34
16.1
47
221
CV
11.7
0.89
0.43
0.54
2.69
% CHG
4.30
0.37
0
0.21
4.70
REJ. COEFF
1.16
0.90
1.13
0.98
0.90
CURRENT
0.97
0.88
1.13
1.03
0.95
STATUS
FAILED OK
OK
OK
OK
Diag. 7.4
If the printer is not used, when the calibration fails, the following menu is displayed:
Rejected and saved coefficients are displayed using the
DOWN
and
UP
keys in order to check
and to record the faulty parameter(s). Press the
ESC
key to return to the main calibration menu.
CALIBRATION FAILED !!!
PRESS A KEY TO CONTINUE ...
SAVED COEFF. WBC 0.97
RBC
0.88
HGB
0.95
REJECT. COEFF.
1.16
0.90
0.90
NOTE:
If the calibration has failed on one or more parameters no parameter is calibrated.
NOTE:
If the calibration fails, the user can restart the calibration, or enter directly the calibration
coefficients through the coefficient function of the calibration menu, using the password.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...