9. MAINTENANCE AND TROUBLESHOOTING
Page 9/17
075D0002-01
Q - "HGB REFERENCE FAILED":
this message is displayed when the instrument gives out of
range HGB blank values after 3 consecutive HGB reference blank cycles.
Check the expiration dates, replace the reagents if necessary or perform a concentrated cleaning.
R - "4 HOURS ELAPSED BETWEEN LAST OPERATION"
: this message is displayed when the
instrument has not been used for at least 4 hours. In order to prevent any drift in the results, a
STARTUP has to be performed before a sample is run. Press the
ENTER
key in order to run the
STARTUP. Run a STANDBY cycle if the instrument has to be stopped.
S - "ERROR : TUBE HOLDER POSITION"
: This message is displayed when the tube holder is
not in its proper position. Turn it slightly to the right or to the left until a click is heard.
T : "ERROR : NO SAMPLE TUBE HOLDER"
: This message is displayed when an analysis
cycle is requested but no sample tube holder is installed.
U : "PLEASE, CLOSE SAMPLE TUBE HOLDER DOOR"
: This message is displayed when the
instrument is set up with the manual start mode and the
START
key is pressed with the tube
holder door open.
Diag. 9.8
V - "TUBE HOLDER DOOR ERROR, PLEASE OPEN
THE DOOR MANUALLY"
: This message is displayed
together with an audible alarm when the tube holder
door is blocked. Open the instrument cover and by
means of a flat screwdriver or a finger nail, push
slightly on the solenoid washer to open the door.
W - "SENSOR ERROR OR DILUENT EMPTY" :
This message is displayed when the instrument
detects a problem on the drainage operation.The waste detection cell may be faulty, or the diluent
may have run out.
X - "MAX. OP. SAVED" :
The operator attempted to change a sixth operator's name on the
current QC card. It is possible to change up to 5 operator names, the sixth one will not be
recorded.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...