![Bayer HealthCare ADVIA 60 Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/bayer-healthcare/advia-60/advia-60_operators-manual_2712837065.webp)
7. CALIBRATION AND QUALITY CONTROL
Page 7/11
075D0002-01
From the MAIN MENU, move the cursor to the function
2
QC and press
ENTER
. The QC menu
is displayed:
QC
1 AUTOMATIC
HH:MM
2 ANALYSIS
7.5. QUALITY CONTROL PROGRAM
7.5.1. Introduction
The
ADVIA 60-
CT
quality control program contains five different functions :
1 - AUTOMATIC
- The function of this submenu is to allow the operator to analyze the commercial
control bloods and store the results on the memory smart card.
2 - ANALYSIS -
The function of this submenu is to allow the operator to analyze a commercial
control with fixed WBC thresholds whatever is the ambient temperature.
3 - PRINT TARGETS
- The function of this submenu is to allow the operator to print the targets
values of the commercial control bloods.
4 - STATISTICS
- The function of this submenu is to allow the operator to print the cumulative
statistics of the commercial control files.
5 - GRAPHS
- The function of this submenu is to allow the operator to print the Levey-Jennings
graphs of the commercial control files.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...