7. CALIBRATION AND QUALITY CONTROL
Page 7/14
075D0002-01
7.5.2.4. Run commercial control
Prepare the commercial control according to the specific instructions (temperature, mixing, etc ... ).
Install the calibrator tube in the correct holder position. Close the sample door to start the analysis
if this mode has been selected or close the sample door and press the cycle button located on
the front panel. When the analysis cycle ends, the first QC results menu is displayed :
7.5.2.5. Accepting/rejecting results
The results are compared to the control assay ranges stored on the memory card. If any result is
out of range, an H (High) or L (Low) will be shown on the display and the printout.
If a third count must be processed, a dollar ($) or a star (*) will be shown on the display and the run
is rejected.
If the HGB blank is not within acceptable limits a (!) is displayed and the run is rejected.
You must rerun the control. The following message appears when the
ENTER
key is pressed:
If the results are correct (no star, dollar or ! will appear), the following menu appears:
The complete listing of results can be rewieved when moving the cursor up or down. Press the
ESC
key to exit this menu.
Diag. 7.7
VALID LOW
NO : ESC
YES : ENTER
RUN REJECTED
PRESS A KEY TO CONTINUE...
WBC
RBC
HGB
HCT
XX.X H
X.XX
XX.X L XX.X
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...