7. CALIBRATION AND QUALITY CONTROL
Page 7/6
075D0002-01
7.3.6. Run calibration
Prepare the calibrator according to the specific instructions (temperature, mixing, etc...).
Press the
ENTER
key. The start calibration menu is displayed :
Check that the results are within the limits given in the calibrator instructions and press
ENTER
.
The first validation menu is displayed:
If the results are not within the acceptable limits, use the
ESC
key to reject the results and to
restart the first run.
If the results are correct, press the
ENTER
key. The menu of the second calibration run is displayed:
Run the second calibrator sample and follow the same procedure until the number of samples set
up is obtained. When the last result has been validated, the instrument calculates the statistical
calibration factors for each parameter : Mean, target, coefficient of variation, percentage difference
between the target value and the mean (% CHG), new calibration coefficient as well as a recall of
the previous calibration coefficient and a pass/fail status appear on the result printout.
Instrument is requesting the first run of the calibrator. Install the calibrator tube in the correct
holder position. Close the sample door to start the analysis if this mode has been selected or
close the sample door and press the cycle button located on the front panel. When the analysis
cycle ends, the first result menu is displayed:
START CALIBRATION
# 2/X
ESC TO EXIT CALIBRATION
ENTER TO ASPIRATE
VALID. CALIBRATION
# 1/X
ESC TO DISCARD
ENTER TO VALID
WBC RBC HGB HCT PLT MPV PRESS ENTER
X.X X.XX XX.X XX XXX X.X
TO CONTINUE
START CALIBRATION
# 1/X
ESC TO EXIT CALIBRATION
ENTER TO ASPIRATE
RUN CAL. ?
NO : ESC
YES : ENTER
NOTE:
Results having flags ($ , * or ! for HGB) are automatically rejected.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...