2. INSTALLATION
Page 2/4
075D0002-01
Diag.2.4
Diag.2.5
Diag.2.6
2.5.2. Connection check
Diag.2.7
Check that the connectors on the printed circuit board
are securely in place.
Re-install the instrument cover.
Push the black plastic carriage locking clip as far as
possible to the left and place the sample needle
carriage as far forward as possible to the right-hand
side. Check that the aspiration needle is not bent
and make sure it is in its upper position.
1 - Carriage locking clip
2 - Sample needle carriage
3 - Needle
Check the position of the chambers. Each chamber
should be in its proper position with its clips and the
electrode block is attached firmly to the RBC
chamber.
1 - Clip
2 - RBC chamber
Unscrew the 2 screws of the WBC/HGB chamber
protection cover slightly. Remove the cover and check
that the chamber is fixed properly in its clips and the
electrode block is attached firmly to the chamber.
Re-install the HGB/WBC chamber cover.
1 - Chamber protection cover
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...