6. STARTUP AND SAMPLE RUN
Page 6/7
075D0002-01
6.4.3. End of the day rinsing
It is necessary to run a standby/shutdown cycle at the end of the day. Press the
STAND BY
key,
the instrument performs a complete cleaning with detergent, and puts the system into the stand
by mode. The instrument can be switched off if the working day is completed or left in this
standby mode overnight or until the next analysis.
6.4. RUNNING SAMPLES
6.4.1. Sample identification
After completing the start up procedure, run a non pathological blood from the previous day,
followed by the quality control analyses and then the sample analyses.
1 - Enter either the patient identification (US mode) or the patient run # (Standard mode).
2 - Install the sample tube in the tube holder and close the sample door to start the analysis if this
mode has been selected or close the sample door and press the
START
key.
6.4.2. Automatic cleaning
When the instrument has run 50 samples from the date changing, an automatic cleaning procedure
is carried out. This cycle lasts 10 minutes approximately while the autocleaning menu is displayed:
AUTO CLEANING
PLEASE WAIT 2MNS 13S
***************************
NOTE:
The automatic cleaning frequency can be adjusted by the user from 1 to 50 as described
in the instrument setup.
When the instrument is left in standby mode, it is necessary to carry out a STARTUP
cycle before any sample analysis.
NOTE:
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...