4. TECHNOLOGY
Page 4/7
075D0002-01
4.3.1. Platelet flags
MIC
following the Platelet result indicates the excessive presence of microcytes in the Platelet
measurement zone. Verify the result using a Platelet Rich Plasma (PRP) or a manual count. This
flag can be adjusted by the user.
SCH
following the Platelet result indicates the presence of schistocytes or Platelet aggregates
in the Platelet measurement zone. Review/scan slide before reporting result. This flag can be
adjusted by the user.
SCL
following the Platelet result indicates the presence of small cells in the 2 and 3 fL zone. A
second sample cycle should be carried out and the results verified. If this flag should persist,
perform an automatic cleaning cycle and resample. If the flag persists verify using a Platelet Rich
Plasma (PRP) of the sample and make a manual slide count for the Platelets. This flag can be
adjusted by the user.
4.3.2. HGB flag
!
: located next to the HGB result shows that the HGB blank carried out during the analysis and
the previous analysis blank differed and were outside the system's precision limits.
Nevertheless the instrument provides a result according to the previous HGB blank. This result
can be reported.
If this suspicious flag occurs more than three consecutive times run the checkup procedure.
4.3.3. WBC Flags
The
ADVIA 60-
CT
has a system of WBC differential flags alerting the operator to the possible
presence of pathological cells, abnormal volume distribution histograms or abnormal elevated
populations such as in the case of the eosinophils and basophils.
1 - Flag
L1
: This flag shows an abnormal number of cells in comparison with lymphocytes (in the
30 - 60 fL zone). The pathologic elements which may be found in this area include platelet
agregates, nucleated red blood cells or atypical lymphocytes. The flag corresponds to the number
of cells counted in the first five channels out of the total number of lymphocytes. The lab operator
can adjust the limit which triggers this flag.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...