9. MAINTENANCE AND TROUBLESHOOTING
Page 9/10
075D0002-01
9.2.3.2. WBC
Check the following if only the WBC count is incorrect or suspicious.
Concentrated cleaning procedure
: Was the concentrated cleaning procedure performed earlier
as part of the identification procedure? If not, perform the concentrated cleaning procedure.
Calibration
: Was the system calibrated earlier as part of the identification procedure. If it was
not, and the WBC count is still erroneous after the concentrated cleaning procedure was performed,
calibrate the instrument as described in section 7. Continue troubleshooting if:
- This does not correct the WBC results.
- The instrument has already been calibrated as part of the identification procedure.
- Earlier attempts to calibrate the WBC during the identification procedure failed.
Analyze a sample and observe the operation of the liquid valve <6>.
Valve <6>
: Is liquid valve <6> opening and closing during analysis cycle? If the liquid valve <6>
is not opening and closing, replace the valve. If this does not correct the WBC count, call your
local technical support provider or distributor.
If the valve is operating properly, call your local technical support provider or distributor.
9.2.3.3. HGB
Check the following if only the HGB results are incorrect or suspicious.
Concentrated cleaning
: Was the concentrated cleaning performed? If it was not, perform the
concentrated cleaning. Continue troubleshooting if this does not correct the HGB result.
Calibration
: Was the system calibrated earlier as part of the identification procedure? If HGB
was not calibrated as part of the identification procedure, calibrate it as described in section 7.
If all attempts to calibrate this parameter have failed, press the
START
key and closely observe the
specific operations of the analyzer listed below in the order specified. Identify the malfunction(s)
and initiate appropriate troubleshooting procedure(s).
If upon close inspection all operations still appear to be acceptable, call your local technical
support provider or distributor.
Lyse pump
: Is the lyse pump plunger moving up and down smoothly? Can you see air bubbles?
If the lyse pump is operating properly and bubbles are not observed, continue. If lyse pump is not
operating properly or bubbles are observed, see Mechanical Checks for troubleshooting procedure.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...