9. MAINTENANCE AND TROUBLESHOOTING
Page 9/9
075D0002-01
Diag. 9.1
Diluent dispenser
: Do you see bubbles in the diluent
dispenser? Is the plunger of the diluent dispenser
moving up and down smoothly during sample
analysis?
Diag. 9.2
If you do not see bubbles and the diluent dispenser
is operating properly, continue. If you see bubbles, or
if the diluent dispenser is not operating properly, refer
to Mechanical Checks for appropriate actions.
If it is, continue. If sample probe is not in its correct
position in the mixing chamber, call your local
technical support provider or distributor.
Diag. 9.3
Transfer of lyse to WBC chamber
: Is the lyse pump
plunger moving up and down smoothly? Can you see
air bubbles?
If the lyse pump is operating properly, and no bubbles
are seen (thus system operations all appear to be
acceptable), call your local technical support provider
or distributor. If bubbles are seen, change the reagent
pack and prime the reagent lines.
Call your local technical support provider or distributor if this does not correct the WBC and HGB
results.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...