9. MAINTENANCE AND TROUBLESHOOTING
Page 9/12
075D0002-01
RBC chamber :
Is the sample probe between the
edge of the chamber and the center of the chamber,
close to the bottom?
Diag. 9.5
9.2.3.5. RBC
Perform the following actions if only the RBC parameter is erroneous or suspicious. Analyze a
normal control and observe the specific operation of the analyzer listed below in the order specified.
Identify the malfunction(s) and initiate appropriate troubleshooting procedure(s). If upon close
inspection all operations still appear to be acceptable, call your local technical support provider or
distributor.
Valve <6>
: Is liquid valve <6> opening and closing during analysis cycle? If the liquid valve <6>
is not operating properly, replace the valve. If the valve is still not operating properly, call your local
technical support provider or distributor. If the valve is operating properly, continue the procedure.
Sampling syringe
: Is the sample syringe moving up and down ? If the 10 µL syringe is not
operating properly during sample analysis, see Mechanical Checks for troubleshooting actions. If
the sampling syringe is operating properly during sample analysis, continue the procedure.
Diluent dispenser
: Do you see any bubbles in the diluent dispenser? Is the plunger of the
diluent dispenser moving up and down smoothly during sample analysis? If no air bubbles are
seen, and the diluent dispenser plunger is operating properly, continue. If bubbles are seen, or the
plunger of the diluent dispenser is not operating properly, see Mechanical Checks for troubleshooting
actions.
Valves <8> and <11>
: Are liquid valves <8> and <11> opening and closing during analysis
cycle? If the liquid valves <8> and <11> are not opening and closing, replace the valves. If this
does not correct the RBC count, call your local technical support provider or distributor
.
If the valves are operating properly, call your local technical support provider or distributor.
If the sample probe is in its proper position in the
RBC chamber (between the edge of the chamber and
the center of the chamber, close to the bottom), and
therefore system operations appear to be acceptable,
call your local technical support provider or distributor.
If the sample probe is not in its proper position in the
RBC chamber (between the edge of the chamber and
the center of the chamber, close to the bottom) call
your local technical support provider or distributor.
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...