7. CALIBRATION AND QUALITY CONTROL
Page 7/3
075D0002-01
7.3.2. Change lot number
The current lot number is displayed. If the lot number of the calibration blood has to be changed,
press
ENTER
. The following menu is displayed:
Without barcode reader:
enter the new lot number (10 characters maximum). Numbers can be
entered directly using the numerical keyboard, letters can be entered using the
DOWN
and
UP
keys. For each letter, press the
ENTER
key to validate it and enter the next one, validate a
second time to record the lot number. The menu turns to the expiration date. If the lot number has
not changed, press the
ESC
key, the menu turns directly to the expiration date.
Diag. 7.2
With barcode reader:
When the calibration blood
tube is equipped with a barcode label, read the
calibration blood identification barcode label using the
barcode reader. When the reading is completed, a
beep occurs and the control lot identification is
displayed on the screen. Press
ENTER
to validate.
The menu turns to the expiration date.
CHANGE LOT # ?
NO : ESC
CURRENT : SETPOINT
YES : ENTER
LOT # :
EXIT : ESC
CURRENT : SETPOINT
SAVE : ENTER
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...