8. INSTRUMENT CONFIGURATION
Page 8/9
075D0002-01
8.3.4. Autocleaning frequency
An automatic cleaning involving a cleaner solution is normally carried out every 50 samples. The
user has the possibilty to setup the automatic cleaning frequency according to the number of
samples usually run in his laboratory. From the SPECIAL menu, move the cursor to the function
4
CLEAN. FREQ. and press
ENTER
. The CLEAN. FREQ.menu is displayed:
Enter the required frequency and press
ENTER
. The new automatic cleaning frequency is recorded.
CLEAN FREQ. ? :
EXIT : ESC
CURRENT :
SAVE : ENTER
8.3.5. Internal setup printout
The setup of all user options can be printed out using this function. Move the cursor to function
5
PRINT CONFIG. and press
ENTER
. The instrument internal setup (limits and configuration) is
printed out.
Diag. 8.2
Содержание ADVIA 60
Страница 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Страница 4: ......
Страница 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Страница 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Страница 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Страница 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Страница 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Страница 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Страница 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Страница 128: ...INDEX Index 2 075D0002 01...