54
Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440
Pressure
Bar
Temperature
°C
Temperature
Min. temp.
Temperature
Max. temp.
0.65
114.51
112.22
116.80
0.70
115.40
113.09
117.71
0.75
116.28
113.95
118.61
0.80
117.14
114.80
119.48
0.85
117.96
115.60
120.32
0.90
118.80
116.42
121.18
0.95
119.63
117.24
122.02
1.00
120.42
118.01
122.83
1.037
121.00
118.58
123.42
1.05
121.21
118.79
123.63
1.10
121.96
119.52
124.40
1.15
122.73
120.28
125.18
1.25
124.18
121.70
126.66
1.30
124.90
122.40
127.40
1.35
125.59
123.08
128.10
1.40
126.28
123.75
128.81
1.45
126.96
124.42
129.50
1.50
127.62
125.07
130.17
1.55
128.26
125.69
130.83
1.60
128.89
126.31
131.47
1.65
129.51
126.92
132.10
1.70
130.13
127.53
132.73
1.75
130.75
128.14
133.37
1.77
131.00
128.38
133.62
1.80
131.37
128.74
134.00
1.85
131.96
129.32
134.60
1.90
132.54
129.90
135.19
1.95
133.13
130.47
135.79
2.00
133.69
131.02
136.36
2.05
134.25
131.57
136.94
2.10
134.82
132.12
137.52
2.15
135.36
132.65
138.07
2.20
135.88
133.16
138.60
2.25
136.43
133.70
139.16
PRESSURE TEMPERATURE CORRELATION CHART
Pressure/Temperature Calibration
This chart shows Pressure and Temperature correlation for Saturated Steam, and the Maximum/Minimum acceptable
Temperature display Values for a Correlation Accuracy of +/- 2%. It applies to Astell sterilisers only.
Note:
This chart is for checking Pressure/Temperature Steam Correlation only and is appropriate for Press/ Temp.
Instruments specified as individually accurate to +/- 1%. It should not be used as a calibration standard for Pressure or
Temperature indicators, which must be calibrated in accordance with the Manufacturer’s Specifications.