Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440
Picture D
Picture E
Picture D
• Push the tool gently into the hole until it engages the
end of the locking bolt.
• Holding the tool, carefully refit the handle loosely onto
the square drive shaft so you can rotate the shaft by
turning the handle.
Picture E
• Push further against the spring-loaded bolt. This will
push the bolt out of engagement.
• Turn the handle clockwise to unlock.
• You can either place the handle loosely on the end of
the square shaft or use a spanner to turn the square
shaft.