51
Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440
STERILISATION PARAMETERS FACTORY SETTINGS
CLASSIC
PARAMETER
Classic
Instrument
Classic
Glassware
Classic
Discard
Classic Fluid
Classic Media
Printer*
ON
FREE STEAMING
Monitor time
60 min
Hold time
5 min
Print interval
1 min
Limit pressure
1200 mbar
Start temperature
95°C
HEAT TO STERILISE
Monitor time
90 min
Print interval
5 min
Stabilize time
1 min
STERILISE
Monitor time
90 min
Print interval
1 min
Sterilisation temp.
121°C
Sterilising time
15 min
Probe chamber**
121=Y
121=Y
121=Y
121=Y
121=Y
Probe load
0=N
121=Y
0=N
121=Y
0=N
121=Y
0=N
121=Y
0=N
121=Y
COOLING
Monitor time
90 min
Print interval
5 min
Cool fan delay
2 min
Cool time
15 min
Cool override time
20 min
Cool temperature
80°C
Cool interlock
Yes
Start temperature
105°C
Cool pressure 1
1200 mbar
Cool pressure 2
1300 mbar
1300
1300
1300
1300
Cool pressure 3
1050 mbar
Cool 1
Yes
Cool 2
No
Hold warm***
No
Temperature
80°C
DRYING
Monitor time
N/A
N/A
N/A
N/A
N/A
Hold time
N/A
N/A
N/A
N/A
N/A
Temperature
N/A
N/A
N/A
N/A
N/A
Cells without entry are the same as in column 2 (classic
instrument).
*
Off if not required, when fitted.
**
Same as sterilise temperature.
***
Requires hold warm to be enabled in settings/system/
door, hold warm.