28
Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440
ALARMS
If any of the functions monitored for faults are triggered
then the system will ensure that a safe condition is
generated.
This may stop the cycle early and then go to cycle
complete failed. The alarm button flashes red.
• Press the alarm button for the alarm list
• Login with your username and password
• Use up and down arrow buttons to select alarm
• When the source of the fault is removed and you press
the acknowledge button, the button changes from
flashing to continuous grey
If the button will not change to grey you have
not cleared the cause of the alarm.
Alarm cancelled
Alarm still active
MANUALLY JUMP STAGES
At any point during the cycle you can press the MENU
button
• Press the stage jump button
• Login with your user name and password as described
earlier
• Use up and down arrow buttons to select the stage you
wish to jump to and press the tick to select
Answer the “are you sure question”
Note:
Alarm icon will flash to show there has been a
manual intervention. You will need to clear this before
completion.
STAGE JUMP AND SECURITY LEVELS
• The stage jump facility is not available to the operator
level
• Astell engineer level can jump all stages
• Stage changes will result in a “failed” cycle