   Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440

•  Touch the screen

•  Touch a blank line on the screen

•  Touch in black box

•  Type the new name

 

•  Press “CONTINUE”

•  Enter the new password for that new user

•  Confirm the new password when asked

•  Press “CONTINUE”

Содержание AMA240

Страница 1: ...MANUAL NUMBER IM_20_80_01J INSTRUCTION MANUAL AMB220 AMB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 AUTOCLAVES STERILIZERS...

Страница 2: ...r logging and audit functions 24 Alarms 28 Manually jump stages 28 Stage jump security levels 28 Safety interlocks 29 Closure system interlocks 30 Heating system 30 Temperature control system 30 Steri...

Страница 3: ...respect and care If correctly used and cared for your steriliser will give you long and safe service Pay attention to regular maintenance Never force the locking mechanism or operate the machine with...

Страница 4: ...01b Survey request added 09 2017 01c IM_20_80_01c Pipework Schematic Updated 01 2018 01d IM_20_80_01d SGS and UKAS logos removed 06 2018 01e IM_20_80_01e Correct part numbers in Parts List 07 2018 01f...

Страница 5: ...voltages Internal voltages are the same range of values as the mains supply to which the unit is connected Hot surface To indicate that the marked item can be hot and should not be touched without ta...

Страница 6: ...nstallation does not comply with the requirements of the installation drawings and other specifications that may be stated by the manufacturers GENERAL INFORMATION Incoming Inspection After receiving...

Страница 7: ...s Faults and malfunctions Service visits Other actions or incidents Check the logbook before you start using the machine as someone else may have recorded a problem or fault of which you are not aware...

Страница 8: ...riliser is operating However depending upon cycle or method of use when the door is open some chamber parts may be hot and could be hazardous to touch Astell cannot advise on specify or be responsible...

Страница 9: ...adly maintained and failed safety valve the user must ensure that all appropriate maintenance is performed to avoid such large steam leaks which are outside the scope of these instructions and outside...

Страница 10: ...downwards to the floor unless specified otherwise at time of order They are in a position to allow access for servicing To cancel warning To cancel run a safety test cycle You cannot cancel it in any...

Страница 11: ...substances Any alternative use or method of operation not covered and specified within this and associated programming manuals is specifically excluded and may be hazardous The manufacturer does not...

Страница 12: ...otate handle to lock 8 Select cycle 9 Press start button cycle will begin 10 When display shows complete at end of cycle press door 11 Door unlocks rotate handle and open door 12 Open door fully and u...

Страница 13: ...ons shown on the screen as rectangles with bevelled corners When a screen shows a button then a press on the button will select that item Screen brightness The Screen contrast is adjustable There is a...

Страница 14: ...at front of machine Accuracy Better than 5 Timing Various timing functions are displayed on the screen in MM SS Instrumentation and system checks In Standby Ready state with door closed the system di...

Страница 15: ...on Display all system alarms Note This button flashes red if an alarm is activated After the alarm icon has been touched the Password is used to cancel alarms Password screen Allows the supervisor to...

Страница 16: ...stage The door will open only if the safety conditions are all satisfied and no alarms are present HANDLE LOCKED WITH DOOR OPEN If the door is open and you move the handle accidentally to the locked...

Страница 17: ...d cycle The display changes to show that a cycle has been selected and loaded Cycle Selected Changes colour to show cycle selected Change cycle If you wish to select another Cycle press the FOLDER but...

Страница 18: ...pe the value you want using the numeric keys then press When you are happy with the date and time shown press the clock button to save the changes When the elapsed time is up the system will go direct...

Страница 19: ...tem will go to the standby stage it may need to cool first this will cause an alarm state Cancel the alarm refer to alarms section you should then be able to restart the sterilize cycle In the case of...

Страница 20: ...ogical assessment is needed before use of the load You will need to judge whether the load has undergone sufficient processing to be sterilised enough to be safe or satisfactory to use Sterilise the l...

Страница 21: ...t security password number ENTERING PASSWORDS Press the menu button for main menu Select settings menu Select edit passwords Now you have to log in with your password Touch your name or if you are a s...

Страница 22: ...ds You cannot use the same password as you used before You cannot use The same password as another user TO CHANGE ADD USER NAME SUPERVISORS SITE ENGINEERS ASTELL ENGINEERS DEVELOPERS The passwords and...

Страница 23: ...MB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 Touch the screen Touch a blank line on the screen Touch in black box Type the new name Press CONTINUE Enter the new password for that new user Confirm t...

Страница 24: ...operator interactions are audited Initial preparations The following information is intended for the person responsible for the administration of the users of the equipment and the maintenance of elec...

Страница 25: ...longer be able to log in Note Only a user with a role of Administrator is permitted to delete another user To Lock a user Select the user to be deleted by pressing on their name in the User Manager s...

Страница 26: ...rd protected pdf copy of the log file by ticking the Export after each cycle shown below If the Export after each cycle is not selected A Printer Archive password protected PDF file will be automatica...

Страница 27: ...opy of the printout from the intergrated printer if this option was selected A password protected Archive file is automatically generated This file contains a record of User login and logout attempts...

Страница 28: ...lashing to continuous grey If the button will not change to grey you have not cleared the cause of the alarm Alarm cancelled Alarm still active MANUALLY JUMP STAGES At any point during the cycle you c...

Страница 29: ...te The temperatures will not be quite the same due to differing thermal inertia between the chamber wall and the load Normally the chamber wall and cooling lock sensor attached to it will cool down fa...

Страница 30: ...ple tubes etc with large numbers of trapped air spaces The temperature at which this starts is just below water boiling point This can be adjusted to allow for altitude effects TEMPERATURE CONTROL SYS...

Страница 31: ...preclude the use in this manner The amount of steam produced when a chamber pressurised to 120 C is exhausted is large and can constitute a serious hazard Unless special high temperature exhaust pipew...

Страница 32: ...ches within 0 5 1cm of the front lip of the chamber Top load models Lift out the load support plate or lower shelf The filling level is correct when the tip of the water level probe dips about 3 mm un...

Страница 33: ...ng Load sensed timing If load sensed process timing is fitted and set up correctly then the system will ensure that the load heats to process temperature as quickly as possible and is processed at the...

Страница 34: ...main sterilisation process The stage timing is controlled by the selected temperature probes and the temperature by the pressure set point The stage continues for set sterilisation time Sterilise tem...

Страница 35: ...main menu settings menu Select edit machine settings Select parameters Select set date and time Login with your password Select the parameter you want to edit Type the value you want using the numeric...

Страница 36: ...e way The container holding the bag must have large holes on the sides to allow steam to enter the container sides and the plastic bag must be slashed or cut with a knife through these holes so that t...

Страница 37: ...r than the chamber temperature and gives up heat to the chamber walls This means that the chamber will reach 80 C before the load so unless load sensed timing is fitted the cooling lock setting will n...

Страница 38: ...pped backwards The usual cause for this is that the water is not entering at the start to the correct level because the chamber water level sensor has been damaged or is missing or is being touched by...

Страница 39: ...positioned the load sensing probe detects the temperature within the load and so will ensure that the load experiences the set conditions for the set time If it is just put to one side of the chamber...

Страница 40: ...ion External user supplied RS232 Please see separate manual for RS232 communications option INTERNAL PRINTER OPTION The standard printer supplied is the Astell type MPP537X This is a 40 character dot...

Страница 41: ...sm Press the larger end with the wheel down gently until the clutch engages then press the other end down until seated correctly Press paper feed for 5 10 seconds to align the ribbon in the slot The r...

Страница 42: ...bon too quickly Printer head does not move Fragments of torn off paper jammed in printer Turn off power Carefully remove with tweezers and turn back on PRINTER SPARE PARTS Part Astell part number Pape...

Страница 43: ...e chlorine bleach cleaners inside the stainless steel chamber Water and detergent should be used to wash the chamber out when required and a scotchbrite scouring pad may be used carefully Do not damag...

Страница 44: ...y should show ready to start Use door button to unlock If the displays show out of sequence are unreliable or do not show or if the machine does not lock correctly turn off and consult a qualified eng...

Страница 45: ...ews holding the d shaped plastic handle retainers to the cover Separate the two d shaped parts Remove handle Pull the handle off the square shaft Note the orientation to make refitting easier This wil...

Страница 46: ...ocking bolt Holding the tool carefully refit the handle loosely onto the square drive shaft so you can rotate the shaft by turning the handle Picture E Push further against the spring loaded bolt This...

Страница 47: ...any cuts or abrasions Check that the gasket is fitted correctly the small holes 4 should be on the outside lip Replace gasket regularly Astell advise that the gasket is a wearing part and will requir...

Страница 48: ...tight and then tighten further with a spanner DO NOT pull on the cable to drag it through the fitting or damage the insulation on the inside of the fitting The sleeved part of the wires should be ins...

Страница 49: ...and air purge times The chamber will heat up to sterilise temperature before the load catches up and this time delay must be accounted for to ensure correct sterilisation The load sensed timing option...

Страница 50: ...plugged in and there is power available at the mains socket If these are all ok call a service engineer Q Difficult to close door against gasket 1 A This could be trapped air expanding If chamber is s...

Страница 51: ...min Probe chamber 121 Y 121 Y 121 Y 121 Y 121 Y Probe load 0 N 121 Y 0 N 121 Y 0 N 121 Y 0 N 121 Y 0 N 121 Y COOLING Monitor time 90 min Print interval 5 min Cool fan delay 2 min Cool time 15 min Coo...

Страница 52: ...ut snap switch 150 C man reset SXQ308 Heater solid state relay PLC670 Door closed micro switch SXS210 Membrane touchscreen surround Consult supplier PT100 sensor assembly Requires also a olive to fit...

Страница 53: ...will certainly need insurance cover and regular inspections Most QA systems require a formal maintenance contract in place and regular calibration must be carried out It is a statutory requirement of...

Страница 54: ...132 73 1 75 130 75 128 14 133 37 1 77 131 00 128 38 133 62 1 80 131 37 128 74 134 00 1 85 131 96 129 32 134 60 1 90 132 54 129 90 135 19 1 95 133 13 130 47 135 79 2 00 133 69 131 02 136 36 2 05 134 25...

Страница 55: ...55 Astell Autoclaves Sterilisers AMB220 AMB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 PRESSURE TEMPERATURE CORRELATION GRAPH...

Страница 56: ...56 Astell Autoclaves Sterilisers AMB220 AMB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 GENERAL FLOW DIAGRAM SCHEMATIC CLASSIC FRONT LOADER...

Страница 57: ...57 Astell Autoclaves Sterilisers AMB220 AMB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 GENERAL FLOW DIAGRAM SCHEMATIC CLASSIC TOP LOADER...

Страница 58: ...58 Astell Autoclaves Sterilisers AMB220 AMB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 GENERAL FLOW DIAGRAM SCHEMATIC AUTOFILL FRONT LOADER...

Страница 59: ...59 Astell Autoclaves Sterilisers AMB220 AMB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 GENERAL FLOW DIAGRAM SCHEMATIC AUTOFILL TOP LOADER...

Страница 60: ...60 Astell Autoclaves Sterilisers AMB220 AMB230 AMB240 AMB420 AMB430 AMB440 AMA240 AMA440 NOTES SPACE This space is for your own notes on the system...

Страница 61: ...Astell Scientific Ltd Powerscroft Road Sidcup Kent DA14 5DT UK Tel 44 0 20 8309 2031 E mail info astell com Astell Scientific Ltd is a company registered in England Registered no 487717 www astell com...

Отзывы: