AX-4030
2-4
AUTION MAX AX-4030 OPERATING MANUAL
B
7
Measure the controls.
1
Press to start.
•
The STAT and control rack will be transported to
the aspirating position while “
Waiting for rack…
” is
displayed.
•
The nozzle will come down to aspirate the first
control, and then the measurement operation takes
place.
•
When the result is obtained, it will be printed. See
“2.8.2 Printed Results Report”
on page 2-50.
•
The following controls are processed in the same
way.
REFERENCE: To view results
Press
Results
to view the latest result. Pressing
<<<<<<<<
or
>>>>>>>>
can display the previous or
next result. To return to the [Measuring…] screen,
press
Go back
.
REFERENCE: To stop control measurement in
progress, press
Stop
. The instrument will
perform the breaking process and put itself on
standby.
Ready for measurement
2006-01-09 13:15
Port No.
0001
Measurement No.
0150
ID
------------------
Test strip type
1: 9EB
Stop
Waiting for rack...
Ready for measurement
2006-01-09 13:17
Port No.
0001
Measurement No.
0150
ID
------------------
Test strip type
1: 9EB
Results Stop
View
Date 2006-01-09 13:17 Kind CONTROL
No. 0001 Po. 001-05 Strip 1: 9EB
ID 000000000000000000
GLU
-
*PRO
4+
BIL
-
URO
NORMAL
PH
7.0
!BLD
+-
*KET
4+
NIT
-
LEU
-
TURB
-
S.G.
1.000
COLOR
LIGHT YELLOW
<<<<<<<< >>>>>>>> Go back