5-6
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
B
An
error
occurs when there is something wrong with the power unit, memory, connections, or
parameter settings. You will be notified of errors by beeps, and can determine the causes by
displayed error codes. The ongoing measurement ceases if an error occurs. You can still obtain the
results of samples that have already been aspirated. However, after remedying the error, you need
to run the tests again on the samples that could not be measured due to the error.
5.2.1
From Error Generation to Remedy
This section describes the procedure from how you will be notified of an error until you clear it.
1
While measurement is in progress, the message
“
Error/trouble occurred
” appears to notify you that
an error (or trouble) has occurred.
•
Sample aspiration will cease.
•
The instrument will wait for the ongoing
measurement to complete if there are test strips
reacting with samples.
2
Consecutive short beeps sound for about 10 seconds
and an error code appears on the screen.
•
If an error occurs while on standby, the instrument
will be brought to in this status immediately.
3
Check the error code and press
OK
to stop the beeping.
4
Take the necessary action to clear the error.
•
See
“5.2.2 Causes and Remedies”
on page 5-7.
5
Run the tests on the samples that could not be measured due to the error.
6
If the error persists, turn off the power and contact your distributor.
REFERENCE: For information on printed lists of error codes reported when an error occurs, see
“2.8.4 Error and Trouble List”
on page 2-54. For information on lists of errors and troubles that
have occurred so far, see
“2.8.5 Trouble List”
on page 2-55 and
“3.5.1 Printing a Trouble List”
on page 3-23.
Measuring...
2006-01-09 11:10
Port No.
0001
Measurement No.
0001
ID
------------------
Test strip type
1: 9EB
Error/trouble occurred
Measuring...
2006-01-09 11:10
Port No.
0001
Measurement No.
0001
ID
------------------
Test strip type
1:10EA
OK
E:102
Error code ("E" and 3-digit number)
Tee-tee-tee, tee-tee-tee
When an Error Occurs
5.2
Содержание AUTION MAX AX-4030
Страница 1: ......
Страница 57: ...1 46 AX 4030 AUTION MAX AX 4030 OPERATING MANUAL ...
Страница 115: ...2 58 AX 4030 AUTION MAX AX 4030 OPERATING MANUAL ...
Страница 181: ...4 38 AX 4030 AUTION MAX AX 4030 OPERATING MANUAL ...
Страница 205: ......
Страница 206: ......