1-24
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
6
Feed the sampler cables through the left side of the
instrument.
7
Loosen the temporarily-tightened screw using a
Phillips screwdriver, and slide the splash guard
down.
NOTE:
Make sure the cables do not get caught in
the splash guard.
8
Tighten the two screws using a Phillips screwdriver to secure the splash guard.
3
Attach the sampler to the instrument.
1
Fit the hooks on both sides of the sampler into the
holes on the instrument.
•
Be careful not to pinch the cables and earth line.
2
Loosen the locking nut on the instrument legs close
to the sampler.
3
Turn the adjuster feet until the sampler rises an even
5 mm off the countertop.
•
Check that the instrument has a clearance of about
20 mm from the countertop.
4
Tighten the locking nuts to lock the adjuster feet in place.
Screw holes
Splash guard
Cables
Hook
Instrument
Sampler
5 mm
About 20 mm
Turn here to adjust height.
Loosen
this nut.
Содержание AUTION MAX AX-4030