AX-4030
4-33
AUTION MAX AX-4030 OPERATING MANUAL
D
4
Remove the old white plate.
1
Slowly pull out the old white plate with a pair of
tweezers.
5
Insert a new white plate.
1
Take a new white plate out of the bag with the
tweezers and, without resting it anywhere, slowly
insert the plate into the slot.
•
Feed the plate through the guides at the entrance
and halfway point, as far as it can go.
IMPORTANT: Do not touch the white plate with
unprotected hands. Touching the plate can
contaminate the surface with sebum, whereby
producing incorrect results.
2
Attach the black sticker to cover the opening.
3
Attach the white plate cover.
•
Ensure the cover is inserted in the slit (between the
two plates).
•
Use a Phillips screwdriver to tighten the screws.
6
Close the cover and turn on the power.
1
Close the front cover.
2
Press the standby switch to turn on the power.
•
Allow the instrument to warm up.
•
The optical block will automatically return to its
home position.
7
Perform check measurement.
1
Measure the check strips to make sure you can obtain correct results.
•
See
“2.7 Check Measurement”
on page 2-43.
White plate
Guides
Viewed from the top
Feed the plate
through the guides
until it stops.
Standby
switch
Front cover