2-22
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
3
Load the sample tubes into the sample racks.
1
Place the sample tubes into the ports of the item
rack.
•
Use an item rack (rack ID #01 to #10) assigned to
the feeder that contains the test strips to be used for
measurement.
•
Optional normal racks can also be used as an
alternative.
NOTE:
If the rack transportation method is set to
“loop”, load the first 0 samples into the start rack.
Starting a batch with another type of rack may
cause double measurement on samples. See page 2-6.
NOTE:
Fit the bottom of the sample tubes into the
recessed portion of the rubber cushions so that
the tubes stand straight. Placing tubes out of
position may cause sample aspiration to fail and
could even damage the nozzle.
REFERENCE: To have the built-in barcode reader
read barcodes successfully, labels on the tubes
must be facing to the rear side of the sample rack.
4
Load the sample racks into the sampler.
1
Hold the sample rack with the marker facing you
and place it on the loading side of the sampler.
•
First, place the rack at the frontward of the loading
side, fit the recesses at the lower right of the rack
into the guides inside the sampler, and slide the rack
backward.
•
The loading side can contain up to 5 racks.
REFERENCE: When the rack transportation
method is set to “loop”, place the start rack at the
beginning. The unloading side of the sampler can
also be used to load racks (up to 5 racks). See page 2-6.
NOTE:
If a sample spills on the sampler, immediately wipe it off. A crystallized sample may
hinder smooth transportation of the sample racks, and consequently cause trouble.
Gray adapters
Item rack
Rubber
cushions
Sample tubes
Barcode label
Rear of sample rack
Marker