4-8
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
4.2.4
Cleaning the Test Strip Stopper
Particles from the test strips can accumulate on the test strip stopper located at the bottom of
the feeders. This may cause running test strips to clog the stopper, consequently leading the
measurement operation to stop with a trouble alert. Clean the test strip stopper every three days
along with the feeders.
Prepare: Supplied blower brush, tissue paper, and protective gloves
IMPORTANT: Wear protective gloves when taking the test strips out of the feeders. Do not
touch the pad area on the test strips with unprotected hands. Test strips contaminated
with sebum may produce incorrect results.
1
Turn off the power.
1
Make sure the standby screen is displayed, then
press the standby switch to turn off the power.
2
Take the test strips out of the feeder.
1
Wearing protective gloves, grasp the test strips in
the feeder, and take them out.
•
For detailed instructions, see step
2
on page 4-6.
2
Wrap the test strips with tissue paper to protect the
strips from dust.
Standby
switch