2-2
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
Preparation for Measurement
2.3
This section describes the tasks you should carry out on a daily basis before testing samples, which
include checking wastes and consumables, turning on the power, assigning the feeders to the item
racks, and preparing test strips and samples.
REFERENCE: Purchase the test strips specified for the instrument because the strips are not
supplied with the instrument.
2.3.1
Checking Wastes and Consumables
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard liquid waste, test strips, and protective gloves in accordance with
local regulations for biohazardous waste.
1
Check for used test strips.
1
Pull out the waste tray to check if there are any used
test strips in it.
•
Discard the used strips, if there are any.
•
See
“4.2.1 Cleaning the Waste Box”
on page 4-3.
NOTE:
Before pushing in the waste box, ensure
the box is correctly sitting on the waste box tray.
Incorrect installation of the waste box may cause
test strips to be scattered inside the instrument or
clog the test strip path.
NOTE:
If test strips are scattered inside the
instrument due to the waste box being detached,
make sure the instrument is on standby, and
remove the strips with tweezers. Wipe the location
where the strips were present using a cloth
moistened with alcohol. Then, install the waste
box correctly.
2
Check for liquid waste.
1
Check if there is any liquid waste in the drain bottle.
•
Discard the liquid waste, if there is any.
•
See
“4.2.2 Discarding Liquid Waste from the Drain
Bottle”
on page 4-5.
Waste box
Waste box tray
Drain bottle