AX-4030
4-5
AUTION MAX AX-4030 OPERATING MANUAL
Replacement of Consumables
4.3
4.3.1
Replacing the Washing Solution
When the instrument is getting low on washing solution, discard the remaining solution and refill
the bottle with newly prepared solution. A maximum of about 600 measurements can be made
with 2 L of washing solution. If warning “W003” occurs due to low washing solution, replace the
solution immediately.
Prepare: Concentrated washing solution 3 (for AUTION MAX), purified water, and sealing film
NOTE:
Do not try to replace the washing solution while the instrument is running. Replacement
while in operation may produce incorrect measurement results or even cause trouble.
1
Remove the washing solution bottle.
1
Make sure the standby screen is displayed.
2
Uncap the washing solution bottle, and discard the
remaining solution.
3
Rinse the bottle with purified water to wash out old
solution.
2
Prepare new washing solution.
1
Dilute concentrated washing solution 3 with
purified water.
NOTE:
Check the correct dilution rate on the bottle
label of the concentrated washing solution 3.
2
Pour the diluted washing solution into the washing
solution bottle.
3
Seal the bottle with a piece of sealing film.
4
While holding the film in your hand, gently invert
the bottle so as not to froth the solution.
Washing solution
bottle cap
Concentrated washing
solution 3
Purified water
1800 mL
200 mL
Example: To prepare 2000 mL of ten-fold
diluted washing solution
Sealing film