Frequency of Maintenance
4.1
The following table lists the maintenance tasks required at directed intervals.
Wear protective gloves to prevent exposure to pathogenic microbes when
performing maintenance tasks marked with a checkmark ( ).
Discard liquid waste, used parts, and cleaning tools in accordance with
local regulations for biohazardous waste.
Alcohol is sometimes used to clean the instrument. Alcohol is readily
combustible, therefore handle it carefully and keep away from flames,
electrical sparks and sources of heat. Also, ventilate the room sufficiently
during use.
Maintenance task
Cycle
See page(s)
Waste box cleaning
Every day or before reaching 400
measurements
4-3
Liquid waste disposal (drain bottle)
Every day or before reaching 600
measurements
4-5
Feeder cleaning
Every three days
4-6
Test strip stopper cleaning
Every three days
4-8
Introduction tray washout
Every three days
4-2
Washing solution replacement
Every 600 measurements
4-5
Thermal printer paper replacement
If red lines appear at both edges of the paper
4-7
Air filter washout
Every month
4-8
Transport tray washout
Every week
4-9
S.G. cell washout
Every week
4-24
Washing bath washout
Every month
4-26
Washing solution filter replacement
Every month
4-29
Drain pinch valve tube replacement
If drainage is leaking from the pinch valve
tube (e.g. “T280”)
4-30
White plate replacement
If warning “W007” occurs
4-32