AX-4030
2-9
AUTION MAX AX-4030 OPERATING MANUAL
2.3.4
Assigning the Type of Test Strips to the Item Racks
Assign a feeder (1, 2, or main) to each item rack (rack ID #01 to #10) used for normal
measurement. This allows the instrument to automatically eject appropriate test strips from the
assigned feeder for measuring samples in the item rack. For example, you can set it so that samples
in the rack #01 are measured with the test strips in feeder 1. Make correct settings to successfully
use the item racks. For more information about the item racks, see
“Sample Racks”
in
“2.1.4
Sample Containers and Sample Racks”
on page 2-7.
NOTE:
In the following procedure, what you actually assign to the item racks are feeders (, 2,
or main), not the type of test strips. If you want to change the type of test strips to load into the
feeder, see the instructions in step
1
in
“2.3.3 Loading Test Strips into the Feeders”
on page
2-6.
1
Access the setup screen.
1
On the standby screen, press
Menu
to go to the
[Main menu]
screen.
2
Press
1
to go to the
[Parameter settings]
screen.
3
Press
1
to go to the
[Rack assignment for test strip]
screen.
Menu Meas No. Strip S.G. Cal
Main menu
(0000) 2006-01-09 10:10
1
Parameter settings
2
Results
3
Initial settings
4
5
Maintenance
Go back
Parameter settings
(1000) 2006-01-09 10:10
1
Rack assignment for test strip
2
Results display format
3
Measurement No. setup