AX-4030
4-5
AUTION MAX AX-4030 OPERATING MANUAL
4.2.2
Discarding Liquid Waste from the Drain Bottle
Liquid waste discharged from the instrument flows out to the external drain bottle. Discard the
liquid waste to empty the bottle before starting the first batch of tests every day. The drain bottle
can contain as much liquid as an amount discharged by about 600 measurements. Check the drain
bottle at times and discard liquid waste before the bottle becomes full.
Prepare: Alcohol and protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard liquid waste and protective gloves in accordance with local
regulations for biohazardous waste.
1
Discard liquid waste.
1
Make sure the standby screen is displayed.
Do not try to discard liquid waste
while the instrument is mechanically
working (e.g., measuring samples).
Uncapping the drain bottle in
operation may splash the liquid
from the tube, causing exposure to
pathogenic microbes.
2
Uncap the drain bottle.
3
Discard liquid waste from the bottle.
2
Sterilize the drain bottle.
1
Sterilize the drain bottle with alcohol.
2
Rinse the bottle with water to remove dirt.
3
Install the drain bottle.
1
Cap the drain bottle, and place it where it was.
Drain bottle
Содержание AUTION MAX AX-4030
Страница 1: ......
Страница 57: ...1 46 AX 4030 AUTION MAX AX 4030 OPERATING MANUAL ...
Страница 115: ...2 58 AX 4030 AUTION MAX AX 4030 OPERATING MANUAL ...
Страница 181: ...4 38 AX 4030 AUTION MAX AX 4030 OPERATING MANUAL ...
Страница 205: ......
Страница 206: ......