2-44
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
B
2
Select the type of measurement.
1
Press the standby switch to turn on the power.
•
Allow the instrument to warm up for about 2
minutes until it puts itself on standby.
2
Press
Menu
to go to the
[Main menu]
screen.
3
Press
5
to go to the
[Maintenance]
screen.
4
Press
3
to go to the
[Check measurement]
screen.
3
Select the feeder for the first check strip (white strip).
1
Select the feeder to load the first white check strip.
•
Press the
hyphen
key to switch between feeders 1
and 2.
2
Press
OK
.
•
You will be asked to load the check strip.
Ready for measurement
2006-01-09 14:00
Measurement start No.
0250
Test strips in feeder 1
1: 9EB
Test strips in feeder 2
2: 9EB
Menu Meas No. Strip S.G. Cal
Main menu
(0000) 2006-01-09 14:00
1
Parameter settings
2
Results
3
Initial settings
4
5
Maintenance
Go back
Maintenance
(5000) 2006-01-09 14:00
1
S.G. cell washout
2
Clean washing bath and tray
3
Check measurement
4
Maintenance information
Check measurement
(5300) 2006-01-09 14:00
Feeder No.
[
1
]
OK Go back