AX-4030
5-
AUTION MAX AX-4030 OPERATING MANUAL
5
Run the tests on the samples that could not be measured due to the trouble.
6
If the trouble persists, turn off the power and contact your distributor.
REFERENCE: For information on the printed list of trouble codes reported when trouble occurs,
see
“2.8.4 Error and Trouble List”
on page 2-54. For information on the printed list of errors and
troubles that have occurred so far, see
“2.8.5 Trouble List”
on page 2-55 and
“3.5.1 Printing a
Trouble List”
on page 3-23.
5.3.2
Causes and Remedies
Wear protective gloves to prevent exposure to pathogenic microbes before
any operation which may expose you to samples.
Discard used samples, test strips and protective gloves in accordance with
local regulations for biohazardous waste.
T999
Unknown trouble
Cause
A system failure occurred.
Remedy Take notes of what appears on the screen, and contact your distributor.
T201
Data ROM trouble
Cause
. The main ROM (version) has been upgraded.
2. Trouble occurred in the flash ROM that stores the parameter settings.
Remedy and 2. Press
OK
to clear the trouble.
↓
Turn off the power, and contact your distributor.
T202
Internal communication trouble
Cause
Internal communication trouble(s) occurred.
Remedy Press
OK
to clear the trouble.
↓
Turn off the power, and contact your distributor.
T203
No calibration curve
Cause
The calibration curve is not set for the test strip, or the calibration curve information is
incomplete.
Remedy Press
OK
to clear the trouble.
↓
Turn off the power, and contact your distributor.