1-18
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
1.4.1
Precautions in Instrument Installation
Before installation of the instrument, read the following notes and always take proper safety
precautions.
■ The instrument weighs about 37 kg and the sampler about 4 kg. Determine a
location for the instrument and assemble it in that location. Do NOT carry the
instrument with the sampler attached. Separate the two units before moving. For
safety reasons, always transport and assemble the instrument with the help of at
least one other person. Hold the bottom of the instrument with both hands when
carrying it.
■ During installation, be careful not to pinch your hands under the instrument.
■ Install the instrument at least 20 cm away from walls. Inadequate clearance
between the instrument and walls may cause overheating of the instrument
or undesirable load on cable connections, thus resulting in fire or incorrect
measurement results.
■ Ensure at least 10 cm clearance between walls and the left panel (viewed from
the front) for users to be able to draw the waste box and dispose of the used
test strips. Insufficient distances may block the waste box. Also, users will have
trouble trying to turn off the main power switch and disconnect connectors in the
event of errors or trouble.
■ Install the instrument where temperature and humidity can be maintained in the
following ranges:
Temperature: 10°C - 30°C
Humidity:
20% - 80%
Installation in the measurement environment outside these ranges may result in
incorrect measurement results.
■ Install the instrument on a level, vibration-free sturdy platform. Operation of
the instrument in an unstable place may cause trouble or malfunction of the
instrument resulting in personal injury. DO NOT install the instrument where it
may fall off or topple over.
■ Do not install the instrument near places that store chemicals or near equipment
that generates corrosive gas or electrical noise. Chemicals, corrosive gases and
electrical noise may cause fire or damage to the instrument and consequently
lead to personal injury, or may otherwise cause incorrect measurement results.
Install the instrument under the supervision of a serviceperson. It is
dangerous to handle the instrument with the covers open. High voltage
parts are located inside. Contact could be dangerous.
Installation
1.4