1-10
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
A
Turbidity Measurement
An internal turbidity/color measurement unit measures sample turbidity using transmitted and
scattered lights.
Light of 635 nm is irradiated on the sample in the cylindrical cell and turbidity is calculated using
the formula below.
T:
Turbidity
Ss:
Scattered light of sample
Ts:
Transmitted light of sample
Sw:
Scattered light of washing solution
Tw:
Transmitted light of washing solution
K:
Factory-set instrument coefficient
Turbidity is determined by the obtained T value.
T range
Turbidity
Output
T < (Level 1)
Clear
–
(Level 1) ≤ T < (Level 2)
Turbid
1+
(Level 2) ≤ T
Densely turbid
2+
* (Level 1) and (Level 2) are constants.
470 nm/525 nm/635 nm
multiple wavelength LED
Cylindrical cell (i.d. 1.8 mm)
Photodiode (transmitted light)
Photodiode (Scattered light)
Ss
Ts
T =
(
– Sw
Tw
)
/
K