AX-4030
5-7
AUTION MAX AX-4030 OPERATING MANUAL
5.2.2
Causes and Remedies
Wear protective gloves to prevent exposure to pathogenic microbes before
any operation that may expose you to samples.
Discard used samples, test strips and protective gloves in accordance with
local regulations for biohazardous waste.
E101
Version change
Cause
The main ROM was exchanged before the instrument was turned on.
Remedy Press
OK
to clear the error.
E102
Power down
Cause
The instrument was accidentally turned off during measurement.
Remedy Press
OK
to clear the error.
↓
Allow the instrument to warm-up. When the instrument puts itself on standby, retry testing the
remaining samples.
E103
Battery voltage error
Cause
The backup battery naturally discharged while the instrument was not in use for several days.
Remedy Press
OK
to clear the error.
↓
Keep the instrument powered on for at least 25 hours to charge the battery.
↓
Set the system clock correctly (see page 3-7).
E104
Backup data error
Cause
The backup data was corrupted.
Remedy Press
OK
to clear the error.
↓
Keep the instrument powered on for at least 25 hours to charge the battery.
↓
Set the system clock correctly (see page 3-7).