4-2
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
4.2.5
Washing the Introduction Tray
Particles from the test strips can accumulate on the introduction tray inside the instrument. Wash
and sterilize the tray thoroughly every three days. This maintenance task can also be done in
the course of transport tray cleaning that should be performed once in a week. For instructions
on cleaning the transport tray at the same time as the introduction tray, see
“4.4.2 Washing the
Transport Tray”
on page 4-19.
Prepare: Alcohol, cloth, and protective gloves, and sponge (soft enough to avoid scratching the
tray)
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used cleaning tools and protective gloves in accordance with local
regulations for biohazardous waste.
NOTE:
Do not use thinner or other organic solvents to wash the introduction trays. Do not
subject the tray to ultrasonic cleaning. Inappropriate cleaning may deform or discolor the tray,
possibly making it unusable.
1
Turn off the power.
1
Make sure the standby screen is displayed, then
press the standby switch to turn off the power.
2
Open the covers.
1
Make sure both feeder covers are locked.
2
Open the maintenance cover.
3
Open the feeder protective cover.
•
The cover stops at a 120° angle.
•
This cover can be removed from the instrument by
unhooking the hinge.
4
Open the side cover.
Standby
switch
Feeder protective cover
Feeder
covers
Side
cover
Maintenance cover
2
3
4