3-22 | IPS-4-UV Analyzer-GPDiv2
a. Press
the
STATUS
soft key to view the
STATUS
screen (Figure 3-5).
Check and record the current operating temperature readings.
This will help you determine if the signal wiring is properly termi-
nated and if all of the circuitry is operating properly.
Pay close attention to the temperatures of the (Oven)
Heater
(if
used), (Optical)
Bench
, and
Cell
(if used) readings. If these tem-
peratures do not increase, check the wiring for proper connec-
tions. If the wiring is properly connected, it may indicate a shorted
RTD for that temperature zone.
The Oven Heater temperature parameter is available only for applica-
tions with a heated Oven.
b. Check the pressure of the pilot valves – Manifold block and Zero
fluid (if used) pressure. Adjust the Actuator/Zero pressure regula-
tor to obtain a minimum pressure of 490 kPag (70 psig / 4.8 barg).
c. Check
the
Aspirator Pressure
(gas applications only) from the
STATUS
screen; the pressure should be adjusted to the value
stated on the system drawings, or enough to see a flow on the
Flow Indicator (or meter), if used, or to achieve an acceptable re-
sponse time. Typically, a good response time is less than 30 seconds
to T90 when the sample system is clean and operating at peak ef-
ficiency – response time may vary due to Sample Line length. The
response is application-specific and will vary, depending on the
sample system.
Figure 3-5.
STATUS menu
(example screen).
> Heater
19 °C
Bench
24 °C
Cell
23 °C
Electronics
24 °C
Manifold Pressure
85 psia
Aspirator Pressure
3 psia
HOME ALARMS
STATUS SETUP
NOTE
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...