5-28 | IPS-4-UV Analyzer-GPDiv2
Alarm Conditions and Corrective Action
This section lists all of the IPS-4 Analyzer system alarm messages (
Condi-
tion/Alarm Name
), describes what triggers the alarms, and provides action
that is required to reset the alarms (
Description & Suggested Corrective Ac-
tion
).
The system alarm relay contact opens on alarm. A system alarm can
result from any of the alarm conditions listed below. Some alarms may
not apply based on the system hardware configuration. For example,
systems without a heated sample system will not have a heater RTD
alarm.
Condition/Alarm Name
Description & Suggested Corrective Action
Heater Temperature
This alarm applies only to systems with a heated Oven/Sample System
Enclosure.
This alarm condition indicates the Oven Heater Plate temperature is
approaching its over-temperature limit; a soft shutdown of the heater
will occur. A soft shutdown means the system will shut down the heater
to ensure the temperature of the heater plate does not continue to
increase to ensure it will not exceed its temperature rating.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
and the Oven/Sample System Enclosure doors, and:
•
Using an Ohm Meter, measure the resistance of the Oven Heater
Plate RTD and test it for an open circuit.
If the RTD is faulty, contact
AMETEK for assistance.
Cell Temperature
Bench Temperature
Oven Temperature
Note:
The Cell and Over Temperature alarms apply only to systems with a
heated Oven/Sample System Enclosure.
These alarm conditions indicate that the specifi c temperature-controlled
zone (Cell, Bench, Oven) is operating
below
or
above
its set point value
by
more
than 5 °C (typical; set point may vary).
If one of these alarms is caused by low temperatures in any of the
temperature-controlled zones, the analyzer will automatically switch to
Not Ready
mode to ensure the system does not become plugged due
to a low temperature in one of its zones.
This error can also be caused by a blown fuse, power interruption, or a
power spike, a failed heater, or wiring problems.
Corrective Action:
For alarms caused by
Low
/
High
Temperatures:
From the User Interface Panel:
- Check current operating temperatures for each temperature zone
and compare them to their set point temperatures.
- Check each of the associated confi guration parameters for
temperature control. These parameters may have been lost as a
result of a power interruption or a power spike.
NOTE
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...