5-18 | IPS-4-UV Analyzer-GPDiv2
10. Perform a leak check on all of the fittings that were reconnected. If
necessary, follow the procedure under “Sample System Leak Check,”
in Chapter 3.
After the leak check has been performed and has passed
, return to this
procedure and continue with the remaining steps.
11.
For applications with heated Sample Systems:
Apply power to the Sample and Vent Line circuitry.
Close
the
Oven Heater
fuse to apply power to the Oven Heater tem-
perature-control zone circuitry.
Close the analyzer enclosure doors and secure them with their latches
to allow the analyzer to warm up to operating temperature and stabi-
lize. This could take up to 2 hours if this sample system is allowed to
cool completely.
12. Zero and Span (if applicable) the analyzer.
Open the Vent Line Shut Off valve (in the Oven/Sample System En-
closure). Open the Sample Line Shut Off valve to allow sample fluid
into the sample system.
13. Switch the analyzer back to
Measure
mode (from the
Calibration
screen, select
Start measure
and press
ENTER
). Press the
HOME
soft key
to return to the
HOME
screen.
14. Adjust the Aspirator Pressure Regulator to achieve normal operating
flow rates, as indicated on system drawings. The Flow Indicator (if
used) should be in the centre of the meter for an optimal flow rate.
15. Close and secure the analyzer enclosure doors with their latches.
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...