5-2 | IPS-4-UV Analyzer-GPDiv2
Maintenance
Generally, there is limited maintenance required to ensure the analyzer
remains operating at peak efficiency, other than that described in the “An-
alyzer Preventive Maintenance Schedule.” This section discusses preven-
tive maintenance that must be followed to ensure continued and proper
operation of the analyzer. This section also describes parts that require
replacing and the frequency in which they should be replaced, according
to this schedule.
Preventive Maintenance
To reduce the occurrence of problems with the analyzer, AMETEK recom-
mends that you follow the “Analyzer Preventive Maintenance Schedule,”
as outlined in the following pages. Since most analyzer problems originate
within the sample handling system, the primary objective of the preven-
tive maintenance schedule is proper care of the sample system.
For Gas Applications:
Plugs in the sample system generally can be avoided by ensuring all
components of the sample system operate at least 10 °C (18 °F) above the
sample dew point temperature. The risk of developing a collection of
liquids in low spots or plugging in the sample system is further reduced
by sloping the Sample Line to the analyzer and the Vent Line back to the
sample stream or other dispersal method.
Preventing leaks in the sample handling system is critical to proper
analyzer operation. The analyzer sample system must be leak checked
whenever it has been dismantled for maintenance or repair. Refer to
“
Sample System Leak Check
” in Chapter 3 for details.
The “Analyzer Preventive Maintenance Schedule” suggests actions and
their frequency when caring for the analyzer. Procedures and exploded-
view drawings that are intended to assist you when changing out parts
can be found in appropriate sections later in this chapter or in a
Manual
Supplement
which is included in the
“Supplemental Information” tab of
this manual and the Documentation Package shipped with the analyzer
.
!
CAUTION
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...