G-2 | IPS-4-UV Analyzer-GPDiv2
118
23 Aspirator pressure high limit
alarm
Alarm
Aspirator pressure (345) is higher
than high limit (372)
118
24 Aspirator pressure low limit
alarm
Alarm
Aspirator pressure (345) is lower
than low limit (373)
118
25 Internal error
Alarm
Hardware and/or software errors that
are internal to the system
118
26 Analog board not present
Alarm
Communication failure to the analog
board
118
27 Discrete board not present
Alarm
Communication failure to the discrete
board
118
28 Discrete board firmware not
compatible
Alarm
Old discrete boards installed in
newer units
118
29 Display board firmware not
compatible
Alarm
Old display boards installed in newer
units
118
30 File system error
Alarm
File system is fragmented; flash chip
is faulty; Unable to read/write to file
system
118
31 Span calibration is required
Warning Detected when span calibration was
never performed prior
118
32 Zero calibration is required
Warning Detected when zero calibration was
never performed prior
119
1 Zero
calibration
failure
Warning Unable to perform zero calibration
119
2 Span 1 calibration failure
Warning Unable to perform span calibration
for analyte 1
119
3 Span 2 calibration failure
Warning Unable to perform zero calibration for
analyte 2
119
4 Span 3 calibration failure
Warning Unable to perform zero calibration for
analyte 3
119
5 Span 4 calibration failure
Warning Unable to perform zero calibration for
analyte 4
119
6 Span 5 calibration failure
Warning Unable to perform zero calibration for
analyte 5
119
7 Span 6 calibration failure
Warning Unable to perform zero calibration for
analyte 6
119
8 Span 7 calibration failure
Warning Unable to perform zero calibration for
analyte 7
119
9 Span 8 calibration failure
Warning Unable to perform zero calibration for
analyte 8
119
10 Dark read failure
Warning Dark pixel readings are not within a
range of pre-defined values
11 Battery failure
Warning System battery fails
119
12 Non-volatile
memory
checksum failure
Warning Non-volatile memory, where system
configuration is stored, does not
return correct checksum value
119
13
119 Warning Zero drift has exceeded a pre-
defined allowable value
119
14 Zero average intensity failure
Warning Zero average readings of the
intensity of the diod array are out of
specification
119
15 Electronic check failure
Warning Verification of the electronic bench
fails
119
16 Verification operation failure
Warning Fail to check verification filter
119
17 Sample temperature high limit
alarm
Alarm
Heater temperature (292) is outside
of a band set by parameter 287
around the setpoint (293)
119
18 Sample Temp low limit alarm
Alarm
Содержание IPS-4-UV
Страница 15: ...Contents xv...
Страница 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 138: ...5 32 IPS 4 UV Analyzer GPDiv2 Figure 5 4 Over Temperature alarm Reset Switch S501 Relay board 100 1942...
Страница 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Страница 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Страница 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Страница 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Страница 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Страница 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Страница 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Страница 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Страница 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Страница 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Страница 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Страница 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Страница 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Страница 200: ...G 4 IPS 4 UV Analyzer GPDiv2 g 1035 17 32 Not used 1036 Not used 1037 Not used 1038 Not used 1039 Not used...